Page 110 Bottle Mode – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
Figure 6-27 Excursion mode setup dialog box
• TOC limit text box allows you to enter a limit an online TOC analysis must reach
before an excursion sample is captured. You may specify a value from 1 to 2000
ppb. The default setting is 500 ppb.
• Run now icon immediately draws a water sample from the online source into the
excursion bottle and opens the bottle filling dialog box. See Figure 6-28.
• Done icon returns you to the home screen. When an excursion condition occurs,
the analyzer performs an excursion sample capture. Following the excursion, the
analyzer always returns to the home screen.
• Cancel icon returns you to the unload/replace bottles dialog box. All changes made
in any of the dialog boxes are lost.
Figure 6-28 Bottle filling dialog box
4) When an excursion occurs, an excursion sample captured dialog box opens, and a
validation occurs. See Figure 6-29.