PAT700 Total Organic Carbon Analyzer – Bottle Mode Page 99
Anatel Operator Manual
Cancel icon takes you to a dialog box that notifies you that the test has been
cancelled. The next button on the cancellation dialog box takes you to the unload/
replace bottles dialog box. All changes made in any of the dialog boxes are lost.
15) When the test running dialog box appears, the analyzer begins running the test.
After verifying the presence of the calibration standard, the analyzer performs a
temperature stabilization. After the temperature has stabilized, the conductivity
standard is analyzed.
The dialog box displays the current status of the test and lists the results of each
repetition and the calibration standards bottle as it is running.
16) When the calibration is completed, the test summary dialog box appears, with the
results of the calibration and a notification of whether the calibration failed or
passed. The results also appear on the printout if a printer is attached. See
Figure 6-17.
17) If the calibration fails, you may run the test again.
Figure 6-17 Test summary dialog box
Summary text lists the results of the calibration. The results appear as they would
appear on the printed report.
Export icon takes you to the export dialog box, where you can save the results of
the calibration to a USB flash drive.
Checkmark icon accepts the calibration, if the values passed, and takes you to the
unload/replace bottles dialog box.
Cancel icon rejects the calibration results and takes you to a dialog box that asks if
you want to rerun the test. If you wish to rerun the test, the analyzer returns you to
the beginning of the test. All previous data are discarded.
NOTE
Immediately upon completion of the test the bottles are backflushed and allowed to
equilibrate to ambient pressure. DO NOT ATTEMPT TO REMOVE THE BOTTLES UNTIL
THE BACKFLUSH AND PRESSURE EQUALIZATION IS COMPLETE.