Page 98 Bottle Mode – PAT700 Total Organic Carbon Analyzer
Operator Manual Anatel
Figure 6-15 Verification resistor installed on printer/calibration port
Figure 6-16 Conductivity meter test dialog box
• Start test icon reads the value of the resistor. The analyzer then converts this
value to conductivity by taking the inverse of the resistance and multiplying it by the
cell constant. This value is then compared to the expected value, assuming a 61.9
Kohm resistor. If the measured value is less than or equal to ±0.1 µS/cm of the
expected value, the test passes and the forward icon is enabled. Otherwise, you are
notified and must cancel the calibration.
Verification resistor