NOTE
Immediately upon completion of the test the bottles are backflushed and allowed to
equilibrate to ambient pressure. DO NOT ATTEMPT TO REMOVE THE BOTTLES UNTIL
THE BACKFLUSH AND PRESSURE EQUALIZATION IS COMPLETE.
11) When the backflush and equalization is completed, the test summary dialog box
appears, with the results of the calibration and a notification of whether the
calibration failed or passed. The results also appear on the printout if a printer is
attached. See Figure 6-11.
Figure 6-11 Test summary dialog box
• Summary text lists the results of the test. The results appear as they would appear
on the printed report.
• Export icon takes you to the export dialog box, where you can save the results of
the test to a USB flash drive.
• Checkmark icon accepts the calibration, if the values passed, and takes you to the
unload/replace bottles dialog box.
• Cancel icon rejects the calibration results and takes you to a dialog box that asks if
you want to rerun the test. If you touch the forward icon, the analyzer returns you to
the beginning of the test. All previous data are discarded.
• Auto accept. If you have selected to auto accept the results and return online, the
results are accepted and sent to the printer, and the analyzer returns to the home
screen. If you have selected manually accept the results and return online, the
analyzer returns to the home screen immediately following the completion of the
test. If you have selected manually accept the results and remain offline, the test
summary dialog box appears.
12) When the unload/replace bottles dialog box appears, follow the instructions for
unloading or replacing bottles, then press the done icon. See Figure 6-12.