Specifications

LASER D22, D23: calibrating the vials
A10
When calibrating the single vial on laser-
transmitter D22, the sweep of the laser can be
used. Place the detector in two positions, at least
1 m apart, and let the laser beam pass through
this from two directions.
Use the three point washers placed on the
magnets for proper elevation, and mount the
laser as shown on picture below.
1. Level according to vial at A.
2. Zero set display at B.
3. Note the value at C.
4. Move the laser to D and level according to vial.
5. Zero set display at E.
6. Note the value at F.
7. Add C- and F-value and divide by 2.
8. Level the laser to the result from point 7.
9. Adjust the vial using a hexagon wrench.
Self calibration of vials when high demands of horizontal
plane. The vials on laser D22 and D23 are normally calibrated
to the laser beam. Measurements that need an absolute
horizontal plane to be the reference will put great demands on
the calibration. Therefore any errors in the calibration are
measured and compensated for. The principle is the same as
for normal calibration, but can get you a better accuracy
because it is done during measurement.
1. Level according to vials.
A. Zero set value.
B. Read the value (for example 1.00)
2. Index the laser 180°, level according to vials.
C. Zero set value.
D. Read the value(for example 2.00)
3. E. Calculate the middle of B and D (in this case 1.50)
This shows the difference in level of the measurement points.
A
B
1.
C
2.
DC
E
B
Calculated value
D
3.