User's Guide

63
8. Troubleshooting
Warning/Error
Screen
Cause
Solution
100%
17-0000
ERROR
Battery too low
Connect to external
power and wait
for the battery
to charge enough
The battery is too low for the meter
to operate, most likely because the
meter has not been used for a very
long time.
Before the meter can be used, you
must connect the meter to an
external power source to charge the
battery enough for it to be used
safely. This may take 30 minutes or
more.