User's Guide

62
Warning/Error
Screen
Cause
Solution
100%
70-21
ERROR
Sample not applied
The sample was not
detected.
Be sure to apply enough
sample when prompted.
The sample was not applied at the
time requested by the meter.
Discard the strip before starting a
new test. Apply the sample when
prompted by the meter.
100%
70-22
ERROR
Meter tilted
The meter must not be
tilted or moved during
the test. Restart the test
with a new strip.
During sample application, the
meter was tilted too much or was
bumped.
Discard the strip before starting a
new test. Do not bump or tilt the
meter after the sample is applied.
100%
70-23
ERROR
LQC lockout
A patient test cannot
be done until a
successful LQC test
has been performed.
Your meter is set to lockout when
either a LQC test fails or a LQC test
is not performed periodically.
A patient test cannot be performed
until the required LQC test(s) is
performed successfully.
Run a passing LQC test (or LQC
tests if your meter requires Level 1
and Level 2 LQC to be performed).
See your administrator for details.
100%
70-24
ERROR
Battery critical
The battery must be
charged. Connect the
analyzer to power before
starting a new test.
The battery is too low to start a
new test.
Before a new test can be started you
must connect the meter to an
external power source, which will
also charge the battery.
100%
70-25
ERROR
LQC expired
The LQC bottle
has expired.
Restart the test
with a LQC kit.
The LQC kit is past its expiry date.
Start the test again with a new LQC
kit that has not expired.