User's Guide

58
Warning/Error
Screen
Cause
Solution
30-02 (Bad fill error)
See error 50-09
See error 50-09
100%
30-03
ERROR
Analysis error
The result could not
be calculated.
Restart the test
with a new strip.
(This row also applies
to 30-04)
An analysis error occurred and the
test result could not be calculated.
Possible causes include too much
meter movement, an unsupported
sample type, a system fault, or test
strips that have not been
maintained properly.
Discard the test strip and start the
test again. Apply the sample as
instructed
on page 19.
If the error persists, contact
Customer Support
(see page 64).
100%
40-01
Below minimum range
The result is below
the minimum range.
INR < 0.8
ERROR
The measured result was below
the measuring range of Xprecia
Prime
(less than 0.8 INR).
Repeat the test with a fresh
fingerstick sample.
Such results should be confirmed
using an alternative test method
(e.g. laboratory PT/INR).
100%
40-02
Above maximum range
The result is above
the maximum range.
INR > 8.0
ERROR
The measured result was above
the measuring range of Xprecia
Prime
(greater than 8.0 INR).
Repeat the test with a fresh
fingerstick sample.
Such results should be confirmed
using an alternative test method
(e.g. laboratory PT/INR).
40-03 (double fill)
See error 50-05
See error 50-05
100%
40-04
Below LQC minimum
The LQC test failed.
See your User Manual
for guidance.
ERROR
The LQC test failed because the
result was below the expected
range.
There are a number of possible
causes for this. See Page 53 for
guidance.
Perform the troubleshooting steps
on Page 53.