User's Guide
56
Warning/Error
Screen
Cause
Solution
Invalid strip barcode
The strip barcode is not
valid. Use only test strips
supplied by Universal
Biosensors.
20-09
WARNING
100%
The meter tried to scan a strip
barcode, but the barcode that was
scanned was not valid.
When prompted to scan a strip,
always scan an Xprecia Prime PT
strip.
You must only use Xprecia Prime PT
strips. All other strips (including
Siemens PT/INR Strips for Xprecia
Systems) will not work.
Invalid vial barcode
The vial barcode is not
valid. Use only test strips
supplied by Universal
Biosensors.
20-10
WARNING
100%
The meter tried to scan a strip vial
barcode, but the barcode that was
scanned was not valid.
When prompted to scan a strip vial,
always scan an Xprecia Prime PT
strip vial.
You must only use Xprecia Prime PT
strips. All other strips (including
Siemens PT/INR Strips for Xprecia
Systems) will not work.
Invalid LQC barcode
The LQC bottle barcode
is not valid. Use only LQC
supplied by Universal
Biosensors.
20-11
WARNING
100%
The meter tried to scan a LQC
bottle barcode, but the barcode
that was scanned was not valid.
When prompted to scan a strip vial,
always scan a Siemens Xprecia
Systems PT Controls bottle.
You must only use Siemens Xprecia
Systems PT Controls. All other LQC
kits will not work.
100%
20-12
WARNING
Analyzer tilted
The analyzer must not be
tilted during the test.
Hold the analyzer level.
The meter was tilted too much or
was bumped.
Hold the meter level to continue the
test. Do not bump or tilt the meter
when the sample is being applied.
100%
20-13
WARNING
LQC test due
A LQC test
should be performed
soon to avoid
a LQC lockout.
Your meter is set to lockout if a
LQC test is not performed
periodically. In the near future, a
lockout will occur unless the
required LQC test(s) is performed
successfully.
You can still run patient tests at this
time, but a LQC test should be
performed soon.
See your administrator for details.