User's Guide
7. Settings
49
When a LQC alert is due, the Test button on the
Home screen will glow yellow.
Enabling LQC Bypass (STAT)
Sometimes a LQC reminder lockout will occur when STAT (Short
Turn Around Time) patient test must be performed. To address this
need, the Supervisor can allow defined operators to bypass a LQC
reminder lockout. Each LQC lockout can only be bypassed up to five
times. See Page 50 for more details.
When LQC is bypassed to perform a patient test,
the test’s historical record will be marked with a
special LQC Bypass padlock icon.
NOTE: bypassing a LQC lockout is only possible if the
lockout is due to a reminder expiring. It is not possible to
bypass a lockout due to a LQC test failing.