Test settings – Display Time
By selecting Tests, then PT and then Result Display Time, the
Supervisor can change how long the test result is shown before the
user is prompted to eject the strip and clean the meter. Swipe up and
down on the Result Display Time screen to increase or decrease the
display time.
Test settings – LQC Lockouts
About LQC Lockouts
The Supervisor has the option of turning on the LQC lockout function.
This function prevents patient tests from being run if:
• the most recently performed LQC test failed or
• a LQC reminder expires.
When a LQC lockout is active, the Test Button on the
Home screen will glow red. If any bypasses are
available (see Page 50), the number of bypasses will
be shown.
Enabling LQC Levels
The Siemens Xprecia™ Systems PT Controls kit contains two
controls focused on different parts of the diagnostic range (low and
high). The Supervisor can enable or disable LQC lockouts altogether.