E. Wait at least two minutes to allow the disinfecting properties of
the CaviWipes™ to take effect.
F. Dry the meter thoroughly with lint free tissues or cloths.
Thoroughly dry the entire surface of the meter being careful to not
allow liquid to enter directly into the strip port connector by
wiping in similar order and direction as shown in the cleaning
sequence.
Pay special attention to places where fluid can gather particularly
around the strip port and the USB port area, drying any liquid that
may have pooled.
G. Put the strip port cover back on the meter having checked that
there is no excess visible liquid especially around the Strip Port.
Ensure that you hear a click sound indicating it is securely re-
attached to the body of the meter.
Signs of deterioration
Contact your authorized distributor (see “Customer Support” on page
64) if you notice any of the following signs of deterioration:
Failure of the device to power on and off.
Fault in button or touchscreen functionality.
The device and button icon do not remain legible.
The display does not remain clear (evidence of cloudiness or fogging)
with cracking or other damage that obscures test results.
Pixels on the display are damaged.
The device casing displays any cracking, swelling, dissolving, softening
or becoming brittle, discolouration.
Decals and labels do not remain legible.