User's Guide
32
Warnings
\
Procedure
Cleaning and disinfecting your meter
Ensure the meter is powered off before cleaning and disinfecting.
A. Start by quickly wiping the whole meter with a CaviWipes™
disposable wipe (damp not dripping) to remove visible debris or
liquids. Pay special attention to the barcode reader window on the
bottom surface, ensuring that this is clear of debris.
Discard the wipe and obtain a fresh wipe before proceeding.
B. Remove the strip port cover using your thumb and forefinger as
shown, then set it aside.
C. Holding the meter level, gently wipe the meter’s front face with
four vertical strokes (twice up and down) starting from the top,
then wipe horizontally in the same cleaning motion (twice left and
right).
Follow the recommended disinfecting instructions and contact times listed
in this User Guide.
Do not use any other disinfectant other than the wipes recommended above
as it may damage the meter.
Squeeze excess liquid out of each wipe before use so that the wipe is damp
but not dripping wet.
Do not let liquid accumulate around the strip port or the strip port cover. If
fluid enters the strip port, your meter might be damaged.
Don’t let liquid pool in the USB port as it may damage the meter.