User's Guide
28
10. Analysis in progress
During analysis, a bar on the bottom of the screen
will show progress and the status of the strip’s
internal quality control check. When the analysis is
complete, a Result screen will appear.
11. Result screen
If the LQC test is successful, the result will be
displayed.
If the LQC test fails, an error screen will be
displayed. If your Supervisor has turned on lock-
outs, you will not be able to perform a patient test
until a successful LQC test is performed. See Page
53 for troubleshooting.
To exit, touch .
12. Eject strip and clean meter
Hold the meter over a waste bin so that the strip
points towards the opening of the waste bin, then
gently press the eject button to discard the test
strip. For disposal instructions, see Page 70.
You will be reminded to clean the meter and strip
port.
You must clean and disinfect the meter and strip
port after each LQC test (see Page 31 for
instructions).
Be careful not to bump the meter or strip
while the sample is being analysed.
If the LQC test fails:
A failed LQC test can indicate that the system is not performing as
expected. It can also be caused by improperly stored controls or an
incorrect test method. Refer to troubleshooting on page 53.