Preparing to run a LQC test
Before running a LQC test you will need to prepare the control
solution using the materials provided in your Siemens Xprecia™
Systems PT Controls kit. To do this:
Open a plasma and diluent bottle
Remove the screw caps from the
plasma bottle and from one of the
diluent bottles.
Add diluent to plasma
Transfer the full contents of the
diluent bottle into the bottle of
plasma using a transfer pipette. Take
care to leave the diluent bottle as
empty as possible. Save the transfer
pipette to use during the LQC test.
Mix the solution
Screw the cap back onto the plasma
bottle firmly and carefully mix the
solution by swirling the vial in a
circular motion until the solution is
completely dissolved. To prevent
foam, do not shake the vial.
Wait at least 5 minutes before using the mixed solution.
You will need to use the solution within 25 minutes (if you keep the lid firmly
closed).
If necessary, you can store the mixed solution for 60 minutes in a
refrigerator at 2–8°C (36–46°F).
Gently swirl the vial again prior to use.