User's Guide

22
Warnings
When running a LQC test, do:
Only use Universal Biosensors Xprecia Prime™ PT Test Strips. Other
test strips (including other Xprecia test strips) will not work with
Xprecia Prime™.
Only use Siemens Xprecia™ Systems PT Controls with the Xprecia
Prime™ meter.
Always follow the safety procedures and precautions listed here and
throughout the user guide, and those adopted by your healthcare
facility.
Always wear protective gloves when performing LQC tests.
Only use the system when the room temperature is between 15 °C
and 32 °C (59 °F to 89 °F) and the relative humidity is less than 80%.
Always store test strips in their original vial and with the cap firmly
closed.
Use a strip within 10 minutes of removing it from the vial.
When you use a test strip vial for the first time, record the discard
date on the vial label in the space provided. This discard date is 3
months after the vial is first opened.
Only use a test strip once and discard after use.
Use the control solutionthat is, plasma that has been reconstituted
with diluentwithin 25 minutes of reconstitution (60 minutes if
refrigerated).
Dispose of hazardous or biologically contaminated materials
according to the practices of your clinic. Discard all materials in a
safe and acceptable manner, and in compliance with all government
requirements.