User's Guide

3. Running a
patient
test
17
2. From the Home screen, tap on the
TEST button
3. Enter Patient ID
If so configured, you will be
prompted for the Patient ID.
Enter the Patient ID by barcode
or using the onscreen
keyboard. You can switch
between barcode/keyboard by
touching the button at the
bottom of the screen.
4. Remove a test strip from its vial
Firmly close the vial immediately after
removing the strip.
Ensure that your gloved hands are clean and dry to
avoid contaminating or damaging the test strips.
Once you have removed the test strip from a vial,
you should conduct the test within 10 minutes.
You can exit a test at any time by touching
the button on the bottom left of the
screen.
To avoid a potentially inaccurate test result, you must use the strip within
10 minutes of removing it from the vial.
Do not use test strips past their expiration date (printed on the label of the
vial) or from a vial that was first opened more than 3 months ago.
10 cm. /
4 inches