User's Guide

14
Warnings
When running a patient test, do:
Only use Universal Biosensors Xprecia Prime™ PT Test Strips. Other
test strips (including other Xprecia test strips) will not work with
Xprecia Prime™.
Always follow the safety procedures and precautions listed here and
throughout the user guide, and those adopted by your healthcare
facility.
Always wear a new pair of protective gloves when testing each
patient.
Always clean and disinfect the meter between each use (for
instructions, see Page 31).
Use only auto-disabling, single-use lancing devices.
Keep the meter as still and level as possible during the test.
Only use the system when the room temperature is between 15 °C
and 32 °C (59 °F to 89 °F) and the relative humidity is less than 80%.
(See “Technical specifications” on page 68 for Transport and Storage
conditions).
Always store test strips in their original vial with the cap firmly snap
closed. Close the strip vial securely (until you hear a “click”) as soon
as you have removed your test strip. This will protect the remaining
strips in the vial.
When you use a test strip vial for the first time, record the discard
date on the vial label in the space provided. This discard date is 3
months after the vial is first opened.
Always store test strips between 2 °C and 30 °C (35.6 °F and 86 °F).
Use a test strip within 10 minutes of removing it from the vial.
Use each test strip only once and discard after use.
Apply the blood sample to the test strip within 30 seconds of sample
collection.
Apply the blood sample to the test strip within 4 minutes of being
prompted to do so by an “apply sample” message on the meter
screen.
Collect only capillary blood (e.g. from a fingerstick) for testing.