User's Guide
3. Running a test
11
Always store test strips in their original vial with the cap firmly snap
closed. Close the strip vial securely (until you hear a “click”) as soon as
you have removed your test strip. This will protect the remaining strips
in the vial.
Always store test strips between 2 °C and 8 °C (35.6 °F and 46.4 °F)
Use a strip within 10 minutes of removing it from the vial.
Use each test strip only once and discard after use.
Try to minimize the time between collecting the sample and applying it to
the test strip.
Do not:
Do not insert a test strip into the strip port more than once. A test strip
that is inserted twice might fail to make a proper electrical contact with
the analyzer.
Do not bump the test strip or analyzer after you apply the sample or while
the test is in progress.
Do not use a vial of test strips if the expiration date recorded on the label
has passed. An error will be displayed if an expired strip is used.
Do not handle a test strip with wet hands because moisture can damage
the strip.
Do not use a strip that has been dropped or which may have been
contaminated.
Do not use a strip if it appears damaged in any way.
Do not shake, stir, aerate or heat a sample prior to testing as these can
all change the measurable free sulfur dioxide levels.
Do not apply the sample to the test strip until you are prompted to do so
by the “apply sample” message on the analyzer screen.
Never add more sample to the test strip after the analysis has begun.