SYMPTOM CAUSE POSSIBLE REMEDY
Whistling • Hearing devices not
inserted properly
• Volume too loud
• Hand/clothing near ear
• Poorly fitting
hearing devices
"Motorboating"
• Low battery
• Dirty battery contact
Not clear, • Poorly fitting
distorted hearing devices
• Hearing devices blocked
with ear wax
• Change in hearing
Weak on the • Volume control set
telephone too low
• Telephone not
positioned properly
• Telephone not telecoil
compatible
Three short • Low battery warning
beeps
For any problems not listed in the guide, contact your hearing healthcare
professional. If you do not have a hearing healthcare professional, please
contact the nearest office listed on the back page of this booklet.
• Remove and reinsert
• Turn volume down
• Remove hand/clothing from
ear
• Consult your hearing health-
care professional
• Replace battery
• Consult your hearing
healthcare professional
• Consult your hearing
healthcare professional
• Refer to section “Cleaning your
Breeze Hearing Instruments”.
Consult your hearing
healthcare professional.
• Consult your hearing
healthcare professional
• Turn up volume
• Move telephone around ear for
clearer signal
• Rent or purchase a telephone
that is telecoil compatible
• Replace battery