8.1.1 GPS communication configuration
Connect to the unit with PQ Online and configure the Computer Port speed to 19200 bps (see Figure
45 and refer also to the PQ Secure User Manual for PQ Online information). This is the maximum
communication rate of the GPS module. This will already be pre-configured if the system is
delivered with the GPS module.
The normal computer port cannot be used when the GPS module is connected. To connect the
instrument to the PC computer, the adapter needs to be removed before the RS-232 cable is inserted
and vice versa.
Figure 45 Computer port configuration
8.1.2 Clock synchronisation configuration
Select the Clock Sync tab and enable the GPS sync. The time zone offset and DST (Daylight Saving
Time) settings are updated when you set the clock in the meter, either from a local PC using
PQOnline or by PQSchedule when the meter is automatically downloaded into a database. For your
information, the current settings are shown in the grey fields.
Press the send button to send the new configuration to the unit and wait at least 2 minutes for all
changes to take effect.
Figure 46 Clock synchronisation configuration
After configuring the meter, set the clock in the meter according to 8.1.3.