Instruction manual
Calibration, Maintenance, Troubleshooting and Warranty
Adjustment Screw Maintenance Procedure
326 MultiMode SPM Instruction Manual Rev. B
15.14.2 Remove Adjustment Screws
To remove adjustment screws, do the following:
1. Remove SPM head and disconnect the scanner body from the Small Sample base by pulling
its cable connector straight up. Hold the scanner body firmly in your hand.
2. Gently turn each screw to check for resistance. Turn counterclockwise until backed out of its
screw hole. If resistance is experienced in turning the screw, stop, rotate briefly in the
opposite direction, then retry. If resistance is experienced on later-model scanners, loosen the
setscrew which applies pressure to the plastic bushing (see diagram above), then try again.
3. If screws are frozen (i.e., cannot be rotated at all), DO NOT attempt to force them. Return the
entire scanner body and screws to Veeco for repair.
15.14.3 Inspect for Physical Damage
Adjustment screws are made of hardened stainless steel; threaded inserts are made of a much softer
brass. If a screw becomes fouled with hard grit it may bind against the threaded insert. If the bound
adjustment screw is forcefully rotated, the screw will almost always destroy the brass insert:
threaded inserts may become cross-threaded, or stripped entirely of threads. If this occurs, they will
have to be pressed out and replaced; return to Veeco for repair.
1. Once screws are removed from the scanner body, they may be washed using methanol. Do
not use strong solvents such as methyl chloride, MEK, benzene, etc. Use a fine brush or swab
to remove grit from between threads and shoulder. Observe caution around rubberized knob
surfaces; certain solvents may dissolve them! Be sure to remove all grit from surfaces; air
dry.
CAUTION: Do NOT splash solvent on the scanner tube or wiring at
the center of the scanner body—certain components
(e.g., wiring insulation) may be dissolved, causing
scanner failure!
ATTENTION: Ne pas éclabousser le tube en céramique piézo-
électrique ou le montage à l’intérieur du tube avec un
solvant. Certains composants pourraient être dissous,
entraînant une défaillance du tube.
VORSICHT: Verwenden Sie keine Lösungsmittel auf dem Scanner-
Röhrchen oder den Kabelanschlüssen am Piezo -
manche Komponenten (z.B. Isolation der
Anschlußdrähte) könnten sich auflösen und eine
Fehlfunktion des Scanners nach sich ziehen.