Instruction manual

Calibration, Maintenance, Troubleshooting and Warranty
Fine-tuning for X-Y Calibration
Rev. B MultiMode SPM Instruction Manual 295
Figure 15.7a Calibration Horizontal Reference
Verify that the microscope’s measured distance agrees with the known horizontal distance. If there
is signicant disagreement between the two, ne tuning is required; go to Step 3 below. If the
displayed distance agrees with the known distance, skip to Section 15.7.3.
3. Based upon the results in Step 2, perform the following calculation:
Divide the known distance by the distance displayed next to the line drawn in Step 2. Record this
value.
4. Select the Realtime > Microscope > Calibrate > Scanner option. The Scanner
Calibration dialog box displays (see Figure 15.7b).
Draw a horizontal line.
Known distance between features
SPM-calculated distance between features
-----------------------------------------------------------------------------------------------------