Instruction manual

Calibration, Maintenance, Troubleshooting and Warranty
272 MultiMode SPM Instruction Manual Rev. B
X-Y Calibration using Capture Calibration and Autocalibration: Section 15.5
Capture Calibration: Section 15.5.1
Autocalibration: Section 15.6
Fine-tuning for X-Y Calibration: Section 15.7
Prepare System for Fine-Tuning: Section 15.7.1
Measure Horizontally at 440V Scan Size: Section 15.7.2
Measure Vertically at 440V Scan Size: Section 15.7.3
Measure Horizontally at 150V Scan Size: Section 15.7.4
Measure Vertically at 150V Scan Size: Section 15.7.5
Change Scan angle and Repeat Calibration Routines: Section 15.7.6
Calibrating Z: Section 15.8
Engage: Section 15.8.1
Capture and Correct an Image: Section 15.8.2
Measure Vertical Features: Section 15.8.3
Correct Z Sensitivity: Section 15.8.4
Recheck Z-axis Measuring Accuracy: Section 15.8.5
Calculate Retracted and Extended Offset Deratings: Section 15.8.6
Finding a Pit with an “A” Scanner: Section 15.8.7
Calibration of “A” Scanners for Atomic-scale Measurement: Section 15.9
Prepare Sample: Section 15.9.1
Quick Guide to Piezo Tube Calibration: Section 15.10
Linearize Scanner: Section 15.10.1
Run Autocalibration Software (Factory Operation): Section 15.10.2
Fine Tune the X-Y Scanner: Section 15.10.3
Calibrate the Z Piezo: Section 15.10.4