Instruction manual
Electric Force (EFM) Imaging
Surface Potential Imaging—Procedure
Rev. B MultiMode SPM Instruction Manual 265
A current-limiting resistor (e.g., 10–100MΩ) should be placed in series with the external voltage
supply to protect the tip and sample from damage. Current-limited power supplies may also be
used. Voltage leads should be connected to pins on the header using soldered, push-on connectors.
Do not solder leads directly to the header pins. Heat may cause damage and/or make jumpering the
pins difficult.
The sample should be electrically connected directly to the chuck or a standard sample puck using
conductive epoxy or silver paint as shown below:
14.7 Surface Potential Imaging—Procedure
1. Locate the two toggle switches on the backside of the Basic Extender box (Figure 14.7a),
then verify that they are toggled as shown in Figure 14.7a.
Figure 14.7a Toggle Switches on Back of Basic Extender Module
Sample Chuck
Sample
Conductive Epoxy
or Paint
FM/Phase
Surface Potential
Gnd/Surface Potential
Analog 2
Mode
Tip or Sample
Voltage
To
NanoScope
To
Microscope