Rev. B MultiMode SPM Instruction Manual 203
11.5.3 High Contact Force
Figure 11.5c shows a curve produced when the tip pushes too far into the sample. The flat portion
on the left side of the amplitude curve in Figure 11.5c occurs because the tip is so close to the
surface that it no longer vibrates. As the piezo extends the tip further, the amplitude of vibration
does not change because the tip is always in contact with the sample surface. The contact force is
very high due to the stiffness of the TappingMode cantilevers. The tip may shatter if deflection
continues.
Figure 11.5c Amplitude Force Plot with High Contact Force
You can avoid this situation by using triggers (see Section 11.3.5) or by reducing the value of Z
scan start until there is no flat portion on the left side of the curve. Rapidly increasing the value of
Z scan start is dangerous because the total oscillation amplitude of the cantilever is small relative
to the total Z travel of the long-range scanner.
11.5.4 Tip Selection
You may use virtually any TappingMode tip to obtain TappingMode force plots; however, the
ultimate choice depends upon the delicacy of the sample and the magnitude of the forces to be
gauged. Longer cantilevers have lower spring constants (i.e., they are more pliant) and therefore
offer greater sensitivity for most samples. Shorter cantilevers afford better control when gauging
strong attractive forces and are less prone to entrapment by surface tension forces. Sharper tips are
more likely to plastically deform the sample and are less sensitive to adhesive forces than tips that
are more blunt. Experiment to determine the tip that best meets your needs.