Instruction manual
TappingMode AFM
Preparation Prior to Imaging
Rev. B MultiMode SPM Instruction Manual 111
Tuning cantilevers manually
The parameter values, especially the drive frequency and the sweep width, given in the following
example apply to one type of cantilever. The nominal parameter values may vary depending upon
the actual cantilever used.
For initial set-up, select View > Sweep > Cantilever Tune. The Sweep Controls panel should be
set to the values shown in the example below (see Figure 7.2f).
Figure 7.2f Sweep control panels for Graph Controls (left) and Main Controls (right).
1. Set parameters in the Sweep Controls panel as follows:
• Set the Drive frequency parameter to a value near the center of the range of the
resonance frequencies specified for the cantilever. For example, if the frequency range is
specified as 240—420kHz, select a drive frequency of 330kHz.
• Start with a Drive amplitude of about 300 mV. It is possible to detach the cantilever
from the substrate by applying too large of a drive amplitude; therefore, it is important
to exercise some caution when adjusting the Drive amplitude parameter. Note that the
Drive amplitude is set automatically during the engage process.
• Set the Sweep width to about the same value as the Drive Frequency, or at least some
value large enough to cover the frequency range specified for the wafer.
• Zero the Amplitude Setpoint. Note that the Amplitude Setpoint will be reset
automatically during the engage process.