Instruction manual

Contact AFM Mode
Optimization of Scanning Parameters
Rev. B MultiMode SPM Instruction Manual 101/(100 Blank)
6.5.4 Setpoint
The Setpoint parameter denes the desired voltage (and, therefore, the desired deection of the
cantilever) for the feedback loop. The setpoint voltage is constantly compared to the present
photodiode cantilever deection voltage to calculate the desired change in the piezo position. When
the gain values are high, as they should be when the Data type is set to Height, the Z piezo position
changes to keep the photodiode output signal close to the Setpoint; therefore, the cantilever
deection remains nearly constant. When the gain values are low as they should be when the Data
type is set to Deection, the piezo height does not change, and the photodiode signal varies around
the Setpoint value.
In Contact Mode, increased Setpoint yields higher tip-sample forces. The Setpoint can be adjusted
to increase or decrease the cantilever deection and, therefore, the contact force of the tip on the
sample. The Force Calibration command in the View menu of Realtime allows the setpoint to be
adjusted while viewing a graph of the tip position versus the deection voltage. Using this
procedure, which is described in detail in the next portion of this section, the contact force of the tip
on the sample can be minimized. This is especially important on soft materials such as biological
samples.
6.5.5 Lowpass lter
The Lowpass lter invokes a digital, one-pole, lowpass lter to remove high-frequency noise from
the Realtime data. The lter operates on the collected digital data regardless of the scan direction.
Settings for this item range from Off through 9. Off implies no lowpass ltering of the data, while
settings of 1 through 9, successively, lower the cut-off frequency of the lter applied to the data
stream.
6.5.6 Highpass lter
The Highpass lter parameter invokes a digital, two-pole, highpass lter which removes low
frequency effects, such as ripples caused by torsional forces on the cantilever when the scan
reverses direction. As with the Lowpass lter, it also operates on the digital data stream regardless
of scan direction. This parameter can be Off or set from 1 through 9. Settings of 1 through 9,
successively, lower the cut-off frequency of the lter applied to the data stream. It is important to
realize that in removing low frequency information from the image, the Highpass lter distorts the
height information in the image. As a result, this lter must be Off when accurate height
information is desired. The Highpass lter is usually used only for atomic images.