Instruction manual

Contact AFM Mode
Suggested Initial Control Settings
94 MultiMode SPM Instruction Manual Rev. B
Feedback Controls Panel
1. Next, select Panels > Feedback.
2. Set both Integral and Proportional gain to 2.0 each and the Deflection Setpoint to 0.0V.
Figure 6.2f Feedback Controls Panel Settings for Initial Setup (Contact Mode)
Other Controls Panel
1. Using the mouse, select Panels > Other to display the Other Controls panel.
2. Verify that the Microscope mode is set to Contact.
3. Set the Z Limit to 440V.
4. Set Units to Volts, Color table to 2.
Figure 6.2g Other Controls Panel Settings for Initial Setup (Contact Mode).