Instruction manual

Contact AFM Mode
Preparation Prior to Imaging
Rev. B MultiMode SPM Instruction Manual 89
Figure 6.1b Laser Adjustment Knobs—Top View
6.1.2 Signal Settings
In Contact AFM Mode, the vertical deection (Vert De.) signal is used to provide the dynamic
feedback signal for surface height tracking. The horizontal deection (Horiz De.) is only used for
lateral force measurements in Contact Mode LFM. When disengaged in Contact AFM Mode and
preparing for engagement, set the detector Vert De. adjustment to between -3.0 and -2.0V for
(SiNi cantilevers) or -0.5 and -1.0 (Si cantilevers). If performing LFM, set Horiz De. to -1.0 to
+1.0V. This is recommended for a 0.0V Setpoint.
Note: Large offsets are not recommended between engage and disengage (2V) with
etched silicon cantilevers in Contact Mode (450µm long only) because
breakage is likely. With etched silicon contact probes, it may be necessary to
reduce the offset between engage and disengage to a value of 1.0—1.5V. The
large difference is recommended for rst-time use of silicon nitride cantilevers.
The difference between the vertical deection before engage and the setpoint is related to the force.
A larger, more positive setpoint voltage results in a larger contact force.
Horizontal Adj.
Vertical.
Laser Spot