User`s guide
4-41
Operation 4
To improve on this basic program, add a wait for
input # 1 timed event at 0.00 min. This means that
the gradient will not start until it receives a signal
from another piece of equipment, for example a
sampler. Only after the sample has been injected
will the gradient start. Program a wait Input # 1 is
closed. To start the program with the wait program-
med, connect Pin 1 to Pin 4 on the Input/Output
connector or press Cancel.
Program List 4
Key to press Notes
Menu
File
3/ENTER File # 3
New/Edit
3/ENTER 3 Loops
4/ENTER Link to file # 4 when finished
Flow Enter the flowrate
0/ENTER At 0.00 minutes
3/ENTER Flowrate = 3mL/min
Add Add the next timed event
Mixt Enter the composition
0/ENTER At 0.00 minutes
5/ENTER 5% solvent B
0/ENTER 0% solvent C
Add Add the next timed event
Mixt Enter the composition
1/ENTER At 1.00 minutes
30/ENTER 30 % solvent B
0/ENTER 0 % solvent C
Add Add the next timed event.
Mixt Enter the composition
2/ENTER At 2.00 minutes
30/ENTER 30 % solvent B
0/ENTER 0 % solvent C
Add Add the next timed event
Mixt Enter the composition
2.5/ENTER At 2.50 minutes
5/ENTER 5 % solvent B
0/ENTER 0 % solvent C
Add Add the text timed event
Mixt Enter the composition
4.0/ENTER At 4.00 minutes
5/ENTER 5 % solvent B
0/ENTER 0 % solvent C
End Finish writing the method
program
Quit
Program List 5
Key to press Notes
Menu
File
4/ENTER File # 4
New/Edit
1/ENTER 1 Loop
None No link at end of file
Mixt Enter the composition
0/ENTER At 0.00 minutes
0/ENTER 0 % solvent B
0/ENTER 0% solvent C
Add Add the next timed event
Flow Enter the flowrate
0/ENTER At 0.00 minutes
0/ENTER 0mL/min
End Finish writing the method
program
Quit Go back to the Ready-to-Run
screen
Programming Examples