User`s guide
4-38
Operation 4
Programming Examples
Programming Examples
Two example programs are given below. Before
running these example programs, ensure that the
hydraulic circuit is properly connected.
Example 1
In the example given in List 1, the flow rate will
start at 1 mL/min and increase to 4 mL/min after
2 minutes. The composition will be constant at
100% solvent A. An output will close at 0.00 min-
utes to start the integrator. The method will only
operate once and will not link to another file when
it is finished. Enter this program following the steps
given Program list 1.
Note that it is necessary to program a flow rate of
0 mL/min at the end of the program to stop the
flow, the flow does not automatically stop when
the program is finished. It will continue with the
last programmed flow rate. To stop the flow at 4.00
minutes you must program a flow rate of 4 mL/min
at 3.99 minutes and 0.00 mL/min. at 4.00 minutes.
Note on Programming Sheet
Appendix G contains a programming sheet which
can be copied. An example of how to use the
programming sheet is also given. The sequence you
should follow is:
• Fill in the File Number and name
• Fill in the SETUP parameters for your system
• Draw the flow rate and composition gradients on
the graphs provided, remembering to mark the
axes
• Fill in the table for Input/Output operations
• Write the programming steps. Finish one section
before going on to the next, i.e. program all of the
flow rate time points before going on to the
composition time points. The 305 software will
arrange all of the time points in the correct sequence.
After writing the method program, it is necessary
to write the programs for the safety/error files.