User`s guide

4-29
Operation 4
Example: Screen 1 3.00 min Output 2
Screen 2 Open Close Pulse
To program this example:
press the Add soft key.
press the Out soft key.
key in the time for the output to operate, 3.00,
and press Enter.
key in the output number, 2, and press Enter.
The display will change to Open, Close or Pulse.
• press the Close soft key.
Inj
In this menu you can injection a sample using the
inject pump. There are three pieces of data for this
menu.
- Time at which the injection pump is started
- Injection volume
- Flowrate for the injection
Example: Screen 1 2.00 min Inj 6.00 mL
Screen 2 Inj rate 0.5 mL/min
To program this example:
press the Add soft key.
press the Inj soft key.
key in the time at which the injection will start,
2.00, and press Enter.
key in the injection volume, 6.00, and press Enter.
key in the value for the injection rate, 0.5, and
press Enter.
The injection rate parameter can be changed to
injection time by pressing the Time soft key. There
is a limit of one injection point for each method
program. The maximum injection volume is the
pump head volume multiplied by 100. If you program
a second injection point, it will write over the
existing injection point.
Preparative injection: The 305 can be used in the
Program mode for repetitive injection of the same
sample.
Programming a Method