User`s guide

4-9
Operation 4
Input/Output Parameter Set Up (I/O)
The I/O menu is used to enter data about param-
eters associated with the complete system. To go to
the Input/Output parameter setup procedure:
press the Menu soft key
press the I/O soft key
The sequence for defining parameters is:
1. High pressure limit.
2. Low pressure limit.
3. Alarm On/Off.
4. GSIOC ID number.
5. Output Contacts.
6. Pause/Program with or without flow.
7. Gradient profile selection.
8. Delay Volume selection.
9. Zero Manometric Module.
The sequence of menu screens and soft key options
is shown on page 4-6.
High Pressure Limit
If the pressure reading from the manometric mod-
ule rises above the defined limit, the pump will
stop. The sequence following a high pressure error
is described page 4-10.
The pressure can be displayed in three different
units; bar, MPa, or kpsi. Change the units dis-
played by pressing the soft key below the units
currently indicated on the display.
The maximum value is the maximum pressure limit
for the manometric module connected in the sys-
tem. The 305 Master pump knows which mano-
metric module is connected and will refuse any
value which is outside the range for that model.
# The default value is the maximum pressure
allowed by the manometric module.
Key in the pressure limit that applies to your
system and press Enter.
Low Pressure Limit
If the pressure reading from the manometric module
drops below the defined limit, the pump will stop.
The sequence following a low pressure error is
described page 4-10. The minimum value is 0.
Setting Up the Pump
This menu will not appear if
there is no manometric module
connected to the system.
Manometric Maximum
Module pressure (bar)
805 600
806 320
807 80