Electrical Installation - Power
Electrical Installation - Power
For safety reasons, the 305 is shipped without the
fuses installed and with the voltage selector in the
220/240 Volt position. You must:
Insert the correct fuses.
Set the voltage selector to your local voltage.
Inserting the Fuses
Ensure that the power cord is not connected before
starting to install the fuses. Follow the procedure
below to install the two fuses.
The voltage selector and fuse holder is located
under neath the power socket. See opposite figure.
Pull the voltage selector out of the power
receptacle. This is done by gently levering the
selector out using a small screwdriver.
Pull out the drawer as shown in the figure
opposite. Insert the first fuse into the clips.
Push the drawer back into position.
Pull out the drawer for the second fuse which is
on the other side of the voltage selector. Insert the
second fuse into the clips.
The instrument requires two fuses to be installed.
The type of fuses required are: 2.0 Amp type “T”
slow blow for 100 -120 V, 1.0 Amp type “T” slow
blow for 220-240 V.
For safety reasons, piston pumps are delivered
without fuses installed. Fuses must be installed by
the user upon delivery.
Selecting the Voltage
The 305 can be set to operate at 100/120 volts or
220/240 volts. The different voltages are selected
depending on the orientation of the fuse holder.
To set the voltage to 100/120 volts:
Insert the fuse holder with the numbers 110/120 on
the bottom, facing the small white arrow.