Technical data

36 Connecting the Smartline Pump 1000 to other Instruments
Table 5 Communication example for the 10 ml pump head
Input Reply Description
ST 2 OK
Set the flow rate to 2 ml/min Command
correctly understood and executed
ST 2.200 OK
Set the flow rate to 2.200 ml/min
Command correctly understood and
executed
ST 10 ?
Set the flow rate to 10 ml/min Command
not understood. Flow rate remains at
2.200 µl/min
Other commands permissible via RS-232 interface to the Smartline pump
1000:
SN Serial number inquiry
GLP GLP inquiry
Date date and time inquiry
VER Program version inquiry
ST Start the pump, using the currently selected flow rate
SP Stops the pump
Pmin minmal pressure inquiry
Pmax maximal pressure inquiry
P Pressure inquiry in Mpa
CONTROL
STANDALONE Enables control from the key pad
CONTROL
REMOTE Enables control from the serial port
If using serial control, the pump sends the following messages to the PC:
OK The last command was successfully executed
E:command The last command could not be understood and/or
executed.