Technical data

Programming the Smartline Pump 1000 23
01 07 001.0 |200| 0 0 0 0 0 0 0 0
OFF 000.00 100 0
Load Edit Prog Flow Pressure Events
Prog Clear 0 [ml/min] [0.1 Mpa] off 0 on 1
2
Hold 0 Time %A%B%C%D
Run 1 [min]
Setup
GLP
Link
Vie
w
A
01 07 001.0 |200| 0 0 0 0 0 0 0 0
OFF 000.00 90 10
Load Edit Prog Flow Pressure Events
Prog Clear 0 [ml/min] [0.1 Mpa] off 0 on 1
2
Hold 0 Time %A%B%C%D
Run 1 [min]
Setup
GLP
Link
Vie
w
B
01 07 - - -.- |200| -- -- -- -- -- -- -- --
OFF
***.*
- - - -
Load Edit Prog Flow Pressure Events
Prog Clear 0 [ml/min] [0.1 Mpa] off 0 on 1
2
Hold 0 Time %A%B%C%D
Run 1 [min]
Setup
GLP
Link
Vie
w
C
01 07
_ _ _._
|200|
_ _ _ _ _ _ _ _
OFF 020.0 10 90
Load Edit Prog Flow Pressure Events
Prog Clear 0 [ml/min] [0.1 Mpa] off 0 on 1
2
Hold 0 Time %A%B%C%D
Run 1 [min]
Setup
GLP
Link
Vie
w
D
Fig. 17 Display of single steps, when creating a program
(see also “Example for creating a program” on page 25)
SOP 14 Execution of Programs
This SOP applies to the Smartline Pump 1000 Firmware Version 1.0X or
higher.
1. Load the program you want to execute by entering the program
number in the LOAD PROG field. Confirm by pressing any arrow key.
2. Press START key to start the pump. The display changes to ON and
the pump begins continuous operation with the parameters specified
in the first program line.
3. Position the cursor onto the ON field.
4. Start the program by pressing the numeric key 1. The display will
show that the program is running by indicating RUN on the lower left
side of the display.
5. While a sequence of programs is being executed, it is possible to
freeze the current status of the pump by pressing the 0 key. The
display will indicate HOLD in the lower left corner and the TIME field
will blink.
6. By pressing again on the 1 key the active program sequence will
continue running again. You can also press STOP to abort execution.