Specifications
33 34
35
38 39 40
The use of Relative Value and Relative 
Percentage Value Mode 
The and % mode applies to all measurement functions, 
with auto ranging.REL mode can be used together with
AUTO-HOLD and MAX/MIN mode.
The definition is as follows:
 Relative value ( )=present value - stored value. 
 For instance, if the stored value is 20.0V and the 
 present value is 22.0V, the reading would be 2.0V. 
 If a new measurement value is equal to the stored 
 value then display 0.0V.
 Relative percentage value ( %) =(relative value 
 stored value) 100%
To enter or exit and % mode:
 Use rotary switch to select the desired measurement 
 function before selecting REL . If measurement 
 functions change manually after REL is selected, 
 the Meter exits the REL  mode.
 Press REL to enter mode, and the present 
 measurement range is locked and display the last 
 measurement value as "0" as the stored value. Then 
 if carrying out measurement at this time, the LCD
 will display the value as the result.
 Press REL  again to enter % mode, and the 
 present measurement range is locked and display 
 the last measurement value as "0" as the stored 
 value. Then if carrying out measurement at this time,
 the LCD will display the value as the % result.
  Press REL  again or turn the rotary switch to reset 
 the stored value and exit and % mode, returning 
 to normal measurement mode.
Turning on the Display Backlight
32
C. Resistance, Conductance & Continuity Test
600Vp
(0.3%+60) 
(0.3%+40) 
 
(0.5%+20) 
(2%+20) 
(2%+120) 
 800.00Ω 0.01Ω  
 8.0000kΩ 0.1Ω  
 80.000kΩ      1Ω  
 800.00kΩ 10Ω  
 8.0000MΩ     100Ω  
 80.000MΩ    1kΩ  
 80.000nS 0.001nS
Continuity Test
0.01Ω
Remarks: 
 Continuity Test Range:
 Buzzer beeps continuously.
 Open circuit voltage approximate 0.7V.
 At 800.00Ω~ 80.000MΩ Range:
 Open circuit voltage approximate 0.7V.
 Conductance Test (80.000nS):
 80nS = 12.5MΩ
D. Capacitance 
 1.000nF 1pF
10.00nF 10pF
100.0nF 100pF
1.000 F 1nF 
10.00 F 10nF 
100.0 F 100nF
At the REL mode
600Vp
E. Diodes Test
  Range Resolution Overload Protection
 Diode 1mV 600 Vp
Remarks:
 Open circuit voltage approximate 3V.
  Displays approximate forward voltage drop reading.
F. DC Current 
 80.000mA 
800.00mA 
8.0000A
10.000A
 1 A 
 10 A 
 100 A 
1mA
 80.000mA 
800.00mA 
8.0000A
10.000A
 1 A 
 10 A 
 100 A 
1mA
(0.2%+40) 
(0.3%+40) 
(0.5%+40) 
(0.8%+40) 
Remarks:
 At 8A & 10A Range: 
 For continuous measurement 10 seconds and 
 interval time between 2 measurements greater
 than 15 minutes. 
G. AC Current 
Remarks:
 At 8A & 10A Range:
 For continuous measurement 10 seconds and 
 interval time between 2 measurement greater than 
 15 minutes. 
37
 Overload
Protection
 1000V
 DC/AC
 rms
 continuous.
800.00mV
80
8.0000V
80.000V
800.00V
1000.0V
100 V
1mV
10mV
100mV
10 V
 (0.8%+60)*
 (0.8%+40)**
 (0.8%+20)
50 ~ 60Hz
(1.5%+60)*
(1.5%+40)**
 (1.5%+20)
40 ~ 50Hz
60 ~ 400Hz
Remarks:
 Input impedance 10M .
 Displays True RMS value.
 Frequency response: 40 ~ 400Hz
 * When input signal 8mV.
 ** When input signal 80mv.
 1000V DC/AC 
 rms 
 scontinuous.
B. DC Voltage 
80.000mV
800.00mV
8.0000V
80.000V
800.00V
1000.0V
1 V
10 V
100 V
1mV 
10mV 
100mV 
 (0.05%+40)
 (0.05%+20)
 (0.05%+20)
 (0.05%+10)
Remarks: 
 At 80.00mV ~ 800.00mV Range: 
 Input impedance 4000M .
  At 8.0000V ~ 1000.0V Range: 
 Input impedance 10M .
4
 Frequency response 40Hz ~ 400Hz.
 Displays True RMS value.
600Vp
2
6
600Vp
6
1A,250V fast type 
glass fuse, 5x20mm. 
10A,250V fast type 
glass fuse, 5x20mm. 
1A,250V fast type 
glass fuse, 5x20mm. 
10A,250V fast type 
glass fuse, 5x20mm. 
1A,250V fast type 
glass fuse, 5x20mm. 
10A,250V fast type 
glass fuse, 5x20mm. 
36
 Maximum Voltage 
 between  
 Terminals and Grounding
5
5
41
80
This meter is suitable for indoor use.
:1A,250V fast type glass 
 fuse, 5x20mm. 
:10A,250V fast type glass 
 fuse, 5x20mm. 
CE Version: IEC61010 
CAT III 1000V, CAT IV 600V 
overvoltage and double
insulation standard.
/UL
UL Version:UL 61010B-1,
CAT II 1000V overvoltage 
double insulation standard.