Specifications
17 18
19 20
21
22 23 24
To avoid harms to you or damages to the Meter from 
electric shock, please do not attempt to measure 
voltages higher than 1000V though the reading may 
be obtained.
The DC voltage ranges are: 8.0000V, 80.000V, 800.00V 
and 1000.0V. To measure DC Voltage, connect the Meter 
as follows: 
When DC voltage measurement has been completed, 
disconnect the connection between the testing leads 
and the circuit under test, and remove the testing 
leads away from the input terminals of the Meter. 
To avoid harms to you or damages to the Meter from 
electric shock, please do not attempt to measure 
voltages higher than 1000V though the reading may 
be obtained.
The DC Millivolt ranges are: 80.000mV, 800.00mV. To
measure DC Voltage, connect the Meter as follows: 
In this range, the Meter has an input impedance 
higher than 4000MΩ
When DC mV voltage measurement has been 
completed, disconnect the connection between 
the testing leads and the circuit under test, and 
remove the testing leads away from the input 
terminals of the Meter.
D. Measuring Resistance / Conductance,
 Capacitance & Continuity 
To avoid damages to the Meter or to the devices 
under test, disconnect circuit power and discharge
all the high-voltage capacitors before measuring 
resistance, capacitance & continuity. Use the DC 
voltage function to confirm that the capacitor is 
discharged.
Resistance / Conductance Measurement
(see figure 6) 
The resistance ranges are: 800.00Ω, 8.0000kΩ,
80.000kΩ, 800.00kΩ, 8.0000MΩ, 80.000MΩ. To 
measure the resistance, Conductance, connect the 
Meter as follows:
1. Insert the red test lead into the terminal
 and the black test lead into the COM terminal.
2. Set the rotary switch to ; resistance 
 measurement is default.
3. Connect the test leads across with the object 
 being measured. The measured value shows on 
 the display.
Note
 The test leads can add 0.1Ω to 0.2Ωof error to 
 resistance measurement. To obtain precision 
  readings in low-resistance measurement, that is the 
  range of 800.00Ω, short-circuit the input terminals 
 beforehand, using the relative measurement function 
 button REL to automatically subtract the value 
 measured when the testing leads are short-circuit 
 from the reading.
 For high-resistance (higher than 1MΩ), it is normal
 that it will take several seconds to obtain a stable 
 reading.
 When measure resistance higher than 12.5MΩ, 
 conductance value can be displayed by pressing 
 HOLD. To switch to resistance value, press HOLD 
 again. 
 Although it is possible to select frequency and duty 
 cycle measurement mode, these functions cannot 
 be used.
 Although it is possible to select frequency and duty 
 cycle measurement mode, these functions cannot 
 be used.
 When the resistance or conductance test has been 
 completed, disconnect the connection between the 
  testing leads and the circuit under test, and remove 
 the testing leads away from the input terminals of 
 the Meter.
Continuity Measurement (see figure 7) 
6
8 9
5
To measure the continuity, connect the Meter as follows:
1. Insert the red test lead into the terminal and 
 the black test lead into the COM terminal.
2. Set the rotary switch to and press to 
 select measurement mode.
3.  The buzzer sounds if the resistance of a circuit under 
 test is less than 100Ω.
 The LCD displays  indicating the circuit being 
 tested is open.
 Press while turning on the Meter to disable the 
 beeper of pressing any button.
 Although it is possible to select frequency and duty 
 cycle measurement mode, these functions cannot 
 be used.
 When continuity testing has been completed, 
 disconnect the connection between the testing leads 
  and the circuit under test. and remove the testing 
 leads away from the input terminals of the Meter.
E. Testing Diodes (see figure 9)
F. DC or AC Current Measurement (see figure 10)
Never attempt an in-circuit current measurement 
where the open-circuit voltage between the circuit 
and ground is greater than 600V.If the fuse burns out 
during measurement, the Meter may be damaged or 
the operator himself may be hurt. Use proper 
terminals, function and range for the measurement.
When the testing leads are connected to the current 
terminals, do not parallel them across any circuit.
 Connect the test leads to the proper terminals as 
 said above to avoid error display. 
 The LCD will display  indicating open-circuit for 
 wrong polarity connection. 
 The unit of diode is Volt (V), displaying the positive
 -connection voltage-drop value.
 Although it is possible to select frequency and duty 
 cycle measurement mode, these functions cannot 
 be used.
 When diode testing has been completed, disconnect
 the connection between the testing leads and the 
 circuit under test.
depending on the resistance of other pathways 
between the probe tips.
To avoid damages to the Meter or to the devices 
under test, disconnect circuit power and discharge 
all the high-voltage capacitors before measuring 
diodes.
Use the diode test to check diodes, transistors, and other 
semiconductor devices. The diode test sends a current 
through the semiconductor junction, then measures the 
voltage drop across the junction. A good silicon junction 
drops between 0.5V and 0.8V.
To test a diode out of a circuit, connect the Meter as 
follows:
1. Insert the red test lead into the terminal 
 and the black test lead into the COM terminal.
2. Set the rotary switch to .
3. For forward voltage drop readings on any 
 semiconductor component, place the 
 terminal (red test lead) to anode & COM terminal
 (black test lead) to cathode.
 The measured value shows on the display.
In a circuit, a good diode should still produce a 
forward voltage drop reading of 0.5V to 0.8V; 
however, the reverse voltage drop reading can vary 
Note
7
Capacitance Measurement (see figure 8) 
Note 
 For testing the capacitor with polarity, connect the 
 red test lead to anode & black test lead to cathode. 
 When capacitance testing has been completed, 
 disconnect the connection between the testing leads 
 and the circuit under test, and remove the testing 
 leads away from the input terminals of the Meter.
The Meter's capacitance ranges are: 1.000nF, 10.00nF, 
100.0nF, 1.000 F, 10.00 F, and 100.0 F. To measure
capacitance, connect the Meter as follows:
1. Insert the red test lead into the terminal and 
 the black test lead or test clip into the COM terminal.
2. Set the rotary switch to and press button 
 to select capacitance measurement mode..
3. Connect the test leads or the test clips across with 
 the object being measured.
 The measured value shows on the display.










