Model UT60F/G: OPERATING MANUAL Table of Contents Page Title Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The multimeter Structure Rotary Switch Functional Button Display Symbols Measurement Ranges Measurement Operation A. DC Voltage Measurement B. AC Voltage Measurement DC or AC Current Measurement C. Measuring Resistance D. Testing Diodes E. Testing for Continuity F. Capacitance Measurement G. H.
Model UT60F/G: OPERATING MANUAL Title Page E. Resistance Diode Test F. Continuity Test G. H. Capacitance Frequency I. Temperature J. Maintenance A. General Service B. Replacing the Battery C.
Model UT60F/G: OPERATING MANUAL Overview This Operating Manual covers information on safety and cautions. Please read the relevant information carefully and observe all the Warnings and Notes strictly. Warning To avoid electric shock or personal injury, read the "Safety Information" and "Rules for Safe Operation" carefully before using the Meter.
Model UT60F/G: OPERATING MANUAL Unpacking Inspection Open the package case and take out the Meter. Check the following items carefully to see any missing or damaged part: Item 1 2 3 4 5 6 Description English Operating Manual Test Lead Test Clip Point Contact Temperature Probe (UT60G only) RS-232C Interface Cable CD-ROM (Installation Guide & Computer Interface Software) Qty 1 piece 1 pair 1 pair 1 piece 1 piece 1 piece In the event you find any missing or damage, please contact your dealer immediately.
Model UT60F/G: OPERATING MANUAL Safety Information This Meter complies with the standards IEC61010: in pollution degree 2, overvoltage category (CAT. III 1000V, CAT. IV 600V) and double insulation. CAT. III: Distribution level, fixed installation, with smaller transient overvoltages than CAT. IV CAT IV: Primary supply level, overhead lines, cable systems etc. Use the Meter only as specified in this operating manual, otherwise the protection provided by the Meter may be impaired.
Model UT60F/G: OPERATING MANUAL Rules For Safe Operation (1) Warning To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter or to the equipment under test, adhere to the following rules: l l l l l l l l l Before using the Meter inspect the case. Do not use the Meter if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic. Pay attention to the insulation around the connectors.
Model UT60F/G: OPERATING MANUAL Rules For Safe Operation (2) l l l l l l l l l Before measuring current, check the Meter’s fuses and turn off power to the circuit before connecting the Meter to the circuit. Replace the battery as soon as the battery indicator appears. With a low battery, the Meter might produce false readings that can lead to electric shock and personal injury.
Model UT60F/G: OPERATING MANUAL International Electrical Symbols AC (Alternating Current). DC (Direct Current). AC or DC. Grounding. Double Insulated. Deficiency of Built-In Battery. Continuity Test. Diode. Fuse. Warning. Refer to the Operating Manual. Conforms to Standards of European Union.
Model UT60F/G: OPERATING MANUAL The multimeter Structure (see figure 1) 5 6 1 2 3 4 ( figure 1) LCD Display Functional Buttons Rotary Switch HzVΩInput Terminal: Input for voltage, frequency, resistance,diode, capacitance and continuity measurements. 5 COM Input Terminal: Return terminal for all measurements. 6 Model UT60F: µA mA Input Terminal; Input for µA and mA measurement. o Model UT60G: mA C Input Terminal Input for mA and temperature measurement. 7 10A Input Terminal Input for A measurement.
Model UT60F/G: OPERATING MANUAL Rotary Switch Below table indicated for information about the rotary switch positions. Rotary Switch Function Position AC or DC voltage measurement V Hz o C µA mA A : Resistance measurement : Diode test : Continuity test Capacitance measurement Frequency measurement Temperature measurement (UT60G only) AC or DC current measurement range from 0.1µA to 4.000mA (UT60F only) Model UT60F: AC or DC current measurement range from 0.01mA to 400.
Model UT60F/G: OPERATING MANUAL Functional Button(1) Below table indicated for information about the functional button operations. Button Measuring Function Operation Performed once to turn the Display Display Press Backlight Backlight on and it shall shut off automatically in around 60 seconds later. V Switches between AC and DC Voltage; the Meter beeps. DC is default. Switches between continuity and diode and resistance measurement; (BLUE) The Meter beeps. Resistance is default.
Model UT60F/G: OPERATING MANUAL Functional Button(2) Any rotary Press HOLD H to enter and exit Switch the Hold mode in any mode, the Position Meter beeps. HOLDH At OFF position l Press and hold HOLD H button while turning on the Meter to display full icons. l Press HOLD H again to return to normal display mode. Any rotary Turn the Meter’s power on and off. Switch Position Any rotary Turn on or off the serial port RS232C switch interface without changing the position original setting.
Model UT60F/G: OPERATING MANUAL Display Symbols(1) (see figure 2) 4 3 5 6 7 16 8 RS232C 1 14 DC 2 AC C F nµmF µmVA Mk Ω Hz 15 9 10/11 12/13 ( figure 2) Number 1 2 3 4 5 Symbol Meaning DC Indicator for DC voltage or current AC Indicator for AC voltage or current The Meter is in the auto range mode in which the Meter autom atically selects the range with the best resolution.
Model UT60F/G: OPERATING MANUAL Display Symbols(2) (see figure 2) 10 11 12 13 14 µA,mA A Amperes (amps). The unit of current. mA: Milliamp, 1x10-3 or 0.001 amperes. µA:Microamp.1x10-6 or 0.000001 amperes. Volts. The unit of voltage. mV: Millivolt. 1x10-3 or 0.001 volts Ω:Ohm. The unit of resistance. Ω,kΩ, kΩ:Kilohm. 1x103 or 1000 ohms MΩ MΩ:Megaohm. 1x106 or 1,000,000 ohms Hertz. The unit of frequency in Hz, kHz, cycles/second. kHz: Kilohertz. 1x103 or MHz 1000 hertz MHz: Megahertz.
Model UT60F/G: OPERATING MANUAL Measurement Ranges(1) A measurement range determines the highest value the Meter can measure. Most Meter functions have more than one range. See “Accuracy Specifications.” A. Selecting a Measurement Range Being in the right measurement range is important: l If the range is too low for the input, the Meter displays OL to indicate an overload. l If the range is too high, the Meter will not display the most accurate measurement. B.
Model UT60F/G: OPERATING MANUAL Measurement Ranges(2) The Meter returns to the autorange mode and displayed. is Note l If you manually change the measurement range after entering the Hold recording modes, the Meter exits these modes.
Model UT60F/G: OPERATING MANUAL Measurement Operation(1) A. DC Voltage Measurement (see figure 3) Select red black ( figure 3) Warning To avoid harms to you or damages to the Meter from eletric shock, please do not attempt to measure voltages higher than 1000V, although readings may be obtained. The DC Voltage ranges are: Model UT60F: 400mV, 4V, 40V, 400V and 1000V; Model UT60G: 600mV, 6V, 60V, 600V and 1000V. To measure DC voltage, connect the Meter as follows: 1.
Model UT60F/G: OPERATING MANUAL Measurement Operation(2) Note l In each range, the Meter has an input impedance of 10MΩ . This loading effect can cause measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10kΩ , the error is negligible (0.1 or less). l When DC voltage measurement has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals. B.
Model UT60F/G: OPERATING MANUAL Measurement Operation(3) 2. Set the rotary switch to V ; and press BLUE button to select AC measurement mode. 3. Connect the test leads across with the object being measured. The measured value shows on the display. Note l In each range, the Meter has an input impedance of 10MΩ. This loading effect can cause measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10kΩ, the error is negligible (0.1 or less).
Model UT60F/G: OPERATING MANUAL Measurement Operation(4) Use the 10A terminal and A range if the current value to be tested is an unknown. Model UT60F has 3 measurement positions on the rotary switch: µA , mA and A . Model UT60G has 2 measuremnet positions on the rotary switch: mA and A . To measure current, do the following: 1. Model UT60F: Insert the red test lead into the µAmA or 10A terminal and the black test lead into the COM terminal.
Model UT60F/G: OPERATING MANUAL Measurement Operation(5) D. Measuring Resistance (see figure 6) Select red black ( figure 6) Warning To avoid harms to you, please do not attempt to measure voltages higher than 60V DC or 30V rms AC. To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring resistance. The resistance ranges are: Model UT60F: 400Ω, 4kΩ, 40kΩ, 400kΩ, 4MΩ, and 40MΩ.
Model UT60F/G: OPERATING MANUAL Measurement Operation(6) Note: l The test leads can add 0.1Ω to 0.2Ω of error to resistance measurement. To obtain precision readings in low-resistance measurement (UT60F: 400Ω and UT60G is 600Ω), short-circuit the input terminals beforehand and record the reading obtained (called this reading as X). Then use the equation: measured resistance value (Y) –(X) = precision readings of resistance.
Model UT60F/G: OPERATING MANUAL Measurement Operation(7) E.Testing Diodes (see figure 7) Select red black ( figure 7) Warning To avoid harms to you, please do not attempt to measure voltages higher than 60V DC or 30V rms AC. To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before testing diodes. Use the diode test to check diodes, transistors, and other semiconductor devices.
Model UT60F/G: OPERATING MANUAL Measurement Operation(8) Note l In a circuit, a good diode should still produce a forward voltage drop reading of 0.5V to 0.8V; however, the reverse voltage drop reading can vary depending on the resistance of other pathways between the probe tips. l Connect the test leads to the proper terminals as said above to avoid error display. l The LCD will display OL indicating either open circuit or wrong polarity connection.
Model UT60F/G: OPERATING MANUAL Measurement Operation(9) Warning To avoid harms to you, please do not attempt to measure voltages higher than 60V DC or 30V rms AC. To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring continuity. To test for continuity, connect the Meter as follows: 1. Insert the red test lead into the HzVΩ terminal and the black test lead into the COM terminal. 2.
Model UT60F/G: OPERATING MANUAL Measurement Operation(10) G. Capacitance Measurement (see figure 9) red black ( figure 9) Warning To avoid harms to you, please do not attempt to measure voltages higher than 60V DC or 30V rms AC. To avoid damage to the Meter or to the equipment under test, disconnect circuit power and discharge all high-voltage capacitors before measuring capacitance. Use the DC votage function to confirm that the capacitor is discharged.
Model UT60F/G: OPERATING MANUAL Measurement Operation(11) Note l For testing the capacitor with polarity, connect the red test lead or red clip to anode & black test lead or black clip to cathode. l To minimize the effect of capacitance stored in the test leads, the test lead should be as short as possible. In addition, the remaining voltage, insulated impedance & dielectric absorption from the capacitor may cause the measurement error.
Model UT60F/G: OPERATING MANUAL Measurement Operation(12) The measurement ranges are: Model UT60F: 4kHz, 40kHz, 400kHz, 4MHz and 40MHz. Model UT60G: 6kHz, 60kHz, 600kHz, 6MHz and 60MHz. To measure frequency, connect the Meter as follows: 1. Insert the red test lead into the HzVΩ terminal and the black test lead into the COM terminal. 2. Set the rotary switch to Hz; 3. Connect the test leads across with the object being measured. The measured value shows on the display.
Model UT60F/G: OPERATING MANUAL Measurement Operation(13) I. Model UT60 G: Temperature Measurement (see figure 11) ( figure 11) Warning To avoid harms to you, please do not attempt to measure voltages higher than 60V DC or 30V rms AC. o o The temperature measurement range is -40 C~1000 C. To measure temperature, connect the Meter as follows: o 1. Insert the red temperature probe into the mA C terminal and the black temperature probe into the COM terminal. o 2. Set the rotary switch to C; 3.
Model UT60F/G: OPERATING MANUAL Sleep Mode To preserve battery life, the Meter automatically turns off if you do not turn the rotary switch or press any button for around 10 minutes. The Meter can be activated by turning the rotary switch or pressing any button except the YELLOW button. The Meter will be under Hold mode and display the last measurement value before it entered sleep mode if it is activated by pressing any button except the YELLOW button.
Model UT60F/G: OPERATING MANUAL General Specifications l Maximum Voltage between any Terminals and Grounding : 1000V rms. l Fused Protection forµAmA Input Terminal : Model UT60F : 0.5A, 250V, fast type fuse, ø5×20mm l Fused Protection for o mA C Input Terminal : Model UT60G : 1A, 250V, fast type fuse, ø5×20mm l Fused Protection for 10A Input Terminal : 10A, 250V, fast type fuse, ø5×20mm Maximum Display : Model UT60F: 3999 Model UT60G: 5999 Auto and manual range selectable.
Model UT60F/G: OPERATING MANUAL Accuracy Specifications(1) Accuracy: (a% reading + b digits), guarantee for 1 year. o o Operating temperature: 18 C~28 C Relative humidity: 75% o Temperature coefficient:0.1×(specified accuracy)/1 C A. DC Voltage Range UT60F UT60G 400.0mV 600.0mV 6.000V 4.000V 60.00V 40.00V 600.0V 400.0V 1000V 1000V Resolution 0.1mV 1mV 10mV 100mV 1V Accuracy (0.5%+5) (0.5%+3) (0.5%+5) Remarks: l Overload protection: 1000V l Input impedance: 10MΩ B.
Model UT60F/G: OPERATING MANUAL Accuracy Specifications(2) C. DC Current Range Resolution UT60F UT60G 400.0µA 0.1µA 4000µA 1µA 40.00mA 60.00mA 10µA 400.0mA 600.0mA 100µA 10.00A 10.00A 10mA Accuracy (1%+2) (1.2%+3) (1.5%+5) Remarks l Overload protection: Model UT60F at 400.0µA to 400.0mA range: 0.5A, 250V, fast type fuse, ø5×20mm Model UT60F at 10.00A range: 10A, 250V, fast type fuse, ø5×20mm Model UT60G at 60.00mA to 600.0mA range: 1A, 250V, fast type fuse, ø5×20mm Model UT60G at 10.
Model UT60F/G: OPERATING MANUAL Accuracy Specifications(3) Model UT60F at 10.00A range: 10A, 250V, fast type fuse, ø5×20mm Model UT60G at 60.00mA to 600.0mA range: 1A, 250V, fast type fuse, ø5×20mm Model UT60G at 10.00A range: 10A, 250V, fast type fuse, ø5×20mm l At 10.00A range: For continuous measurement 10 seconds and interval time between 2 measurements greater than 15 minutes. l Frequency response: 40Hz~1kHz. E. Resistance Range ResolutionAccuracy Remarks UT60F UT60G Open circuit voltage 0.1Ω (1.
Model UT60F/G: OPERATING MANUAL Accuracy Specifications(4) H. Capacitance Range Overload ResolutionAccuracy Protection UT60F UT60G (4%+10) 4.000nF 6.000nF 1pF 40.00nF 60.00nF 10pF 400.0nF 600.0nF 100pF (4%+3) 4.000µF 6.000µF 1nF 600Vp 40.00µf 60.00µF 10nF 400.0µF 600.0µF 100nF (5%+10) 4.000mF 6.000mF 1µF I. Frequency Range Resolution UT60F UT60G 1Hz 4.000kHz 6.000kHz 10Hz 40.00kHz 60.00kHz 100Hz 400.0kHz 600.0kHz 1kHz 4.000MHz 6.000MHz 10kHz 40.00MHz 60.00MHz Accuracy (0.
Model UT60F/G: OPERATING MANUAL Maintenance(1) This section provides basic maintenance information including battery and fuse replacement instruction. Warning Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information. To avoid electrical shock or damage to the Meter, do not get water inside the case. A. General Service l l l l l Periodically wipe the case with a damp cloth and mild detergent.
Model UT60F/G: OPERATING MANUAL Maintenance(2) Warning To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator “ ” appears. Make sure the test leads are disconnected from the circuit being tested before opening the case bottom. To replace the battery: l Press the POWER to turn the Meter off and remove all connections from the terminals.
Model UT60F/G: OPERATING MANUAL Maintenance(3) Warning To avoid electrical shock or arc blast, or personal injury or damage to the Meter, use specified fuses ONLY in accordance with the following procedure. To replace the Meter’s fuse: l Press the POWER to turn the Meter off and remove all connections from the terminals. l Remove the screw from the battery compartment, and separate the battery compartment from the case bottom.
Model UT60F/G: OPERATING MANUAL RS232C Serial Port(1) A. RS232C Port Cable The Meter D-sub 9 Pin Male Computer D-sub D-sub 9 Pin 25 Pin Female Female Pin Name 2 2 3 RX 3 3 2 TX 4 4 20 DTR 5 5 7 GND 6 6 6 DSR 7 7 4 RTS 8 8 5 CTS B. Setting of RS232C Serial Ports Default of RS232C Serial Port for communication is set as: Baud Rate 19230 Start bit 1 (always 0) Stop bit 1 (always 1) Data bits 7 Parity 1 (odd) C.
Model UT60F/G: OPERATING MANUAL RS232C Serial Port(2) l l l Can access to a local or a network CD-ROM A free serial port A mouse or other pointing device supported by Windows. Please refer to the included CD-ROM “Installation Guide & Computer Interface Software” for installing and operating instructions of the Interface Program. ** END ** This operating manual is subject to change without notice.
Model UT60F/G: OPERATING MANUAL 41
Model UT60F/G: OPERATING MANUAL Copyright 2001 Uni-Trend International Limited. All rights reserved. Manufacturer: UNI-TREND TECHNOLOGY(DONG GUAN)LIMITED Address: Dong Fang Da Dao, Bei Shan Dong Fang Industrial Development District, Hu Men Town, Dong Guan City, Guang Dong Province, China Headquarters: Uni-Trend International Limited Address: Rm901, 9/F, Nanyang Plaza 57 Hung To Road Kwun Tong Kowloon, Hong Kong Tel: (852) 2950 9168 Fax: (852) 2950 9303 Email: info@uni-trend.com http://www.uni-trend.