User Manual

Select waveform of Burst
After setting N-cycle mode of pulse train, press Carrier Type Sine to select carrier
signal as sine wave. The default waveform of pulse train is sine wave, so it is unnecessary
to change in this example.
Figure 4-53 Select waveform of pulse train
You can set amplitude with multi-functional knob and direction key (note: if frequency is
displayed, only frequency can be set, which means that conversion between frequency
and period cannot be realized. If frequency is displayed, period of 2ms is corresponding
to frequency of 500Hz. They are reciprocal, i.e. T=1/f). You can also press soft function
key Param Freq → Freq again (press the soft key Freq twice for conversion between
frequency and period in parameter list), when the interface below will pop up. To set some
parameter, press corresponding soft key, input the required value and select the unit.
Figure 4-54 Set waveform amplitude