User Manual
b)Set Starting and Stop Frequency
Starting frequency and stop frequency are upper limit and lower limit of frequency sweep.
Function/ arbitrary waveform generator always sweeps from starting frequency to stop
frequency and then returns to starting frequency. To set starting or stop frequency, please
press Param → StartFreq or StopFreq , input number through numeric keyboard and press
corresponding soft key of unit to finish setting.
Figure 4- 33 Set frequency sweep parameters
When starting frequency < stop frequency, function / arbitrary waveform generator sweeps
from low frequency to high frequency.
When starting frequency> stop frequency, function / arbitrary waveform generator sweeps
from high frequency to low frequency.
When starting frequency= stop frequency, function / arbitrary waveform generator outputs
fixed frequency.
The synchronous signal of frequency sweep mode is low from starting point to midpoint
of frequency sweep time, and high from midpoint to end of frequency sweep time.
By default, starting frequency is 1kHz and stop frequency is 2kHz, but the range of starting
and stop frequency can vary with frequency sweep waveform. See the table below for the
frequency range of frequency sweep wave:
Table 4- 4
UTG2122B UTG2082B UTG2062B
1Hzµ
1Hzµ
1Hzµ
1Hzµ
1Hzµ
1Hzµ
1Hzµ
1Hzµ
1Hzµ
1Hzµ
Frequency
Carrier
waveform
Sine
Square
Ramp
Pulse
Arbitrary
Min MinMin
1Hzµ
1Hzµ
1Hzµ
1Hzµ
1Hzµ
Max Max Max
120MHz
80MHz
5MHz
30MHz
25MHz
80MHz
70MHz
4MHz
25MHz
20MHz
60MHz
60MHz
3MHz
20MHz
15MHz