After you use QAM function, you can set QAM rate, which is in the range of 2mHz~1MHz
and 100Hz by default. You can change it with multi-functional knob and direction key on the
interface for using ASK function or by pressing Param Rate
Comprehensive example
First make the instrument run in QAM mode, and then set an internal sine wave of 2kHz
and 2Vpp as carrier signal. Set rate to be 100Hz and modulation mode to be 64QAM,
default source is PN7. The specific steps are as follows:
1)QAM function
Press Mod Type QAM successively (press soft key Type to select if Type is not
highlighted) to use QAM function.
Figure 4-14 Select QAM function
2)Set carrier signal parameters
Press the key of basic waveform type setting Sine to select carrier signal as sine wave.
The default carrier signal is sine wave, so it is unnecessary to change in this example.
You can set by pressing Carrier Type Sine . You can also press corresponding soft
keys of function again, when the interface below will pop up.