User Manual

Set modulation depth
Modulation depth is degree of amplitude change, expressed in percentage. The range of
SUM depth is 0%~100%, 100% by default. When modulation depth is 0%, carrier wave is
output. When modulation depth is 100%, modulation wave is output. You can change it with
multi-functional knob and direction key on interface for using amplitude modulation function
or by pressing Depth . In case of external modulation source, the output amplitude of
instrument is also controlled by±5V signal level on external analog modulation input terminal
(Modulation In connector) of back panel.
Comprehensive example
First make the instrument run in SUM mode, and then set an internal sine wave of 1kHz as
modulation signal and a square wave with frequency of 2kHz, amplitude of 200mVpp and
duty ratio of 45% as carrier signal. Finally set modulation depth to be 80%. The specific
steps are as follows:
1)Use SUM function
Press Menu Mod Type SUM (press soft key Type to select if Type is not
highlighted).
Figure 4- 7 Select SUM function
2)Select carrier waveform
After enabling SUM function, press Carrier Type Square .
You can press Param again until the following interface appears.
60