User Manual

Hop Frequency Setting
After enabling FSK function, the default of hop frequency is 2MHz. If need to change, press
Param HopFreq in turn. Settable range of hop frequency is determined by carrier wave
waveform. See Carrier Wave Frequency Setting
FSK Rate Setting
When modulation source is internal, the moving frequency between carrier wave frequency
and hop frequency can be set. After enabling FSK function, FSK rate can be set and the
settable range is 2mHz to 1MHz, the default rate is 100Hz. If need to change, press
Param Rate in turn. Please refer to Carrier Wave Frequency to set up the frequency.
Comprehensive Example
Firstly, make the instrument work in frequency shift keying (FSK) mode, then set a sine
wave with 2kHz and 1Vpp from the internal of the instrument as a carrier wave signal, and
set hop frequency to 800 Hz, finally, make carrier wave frequency and hop frequency move
between each other with 200Hz frequency. Specific steps are seen as following:
Enable Frequency Shift Keying (FSK) Function
Press Menu Mod Type FSK in turn to start the FSK function.
Set Carrier Wave Signal Parameter
Press Carrier Type Sine in turn to select sine wave as carrier wave.
40