User Manual

e)Set pulse wave
Example: period=2ms, amplitude=1.5Vpp, offset=0V, duty=25%, rising time=200us and
falling time=200us:
1. Press Menu Wave Type Pulse Param , and then press Freq to switch to Period.
Enter required number value and select the unit. When entering duty ratio value, there is
a label at the bottom of display, and select 25%.
2. Press Param to enter sub menu to set falling edge time, then press Fall to enter required
number and select unit.
Note: This parameter can be set by multifunctional knob and direction buttons.
f)Set DC offset voltage
Default DC voltage: 0V.
Steps for setting DC offset voltage to 3V:
1.Press Menu Wave Type DC . If the current DC voltage is valid, adapt the same
DC voltage value.
2.Enter 3.
3.Select required unit V
Note: This parameter can be set by multifunctional knob and direction buttons.
11