User Manual

c) Set DC offset voltage
Default waveform: sine wave with 0V DC offset voltage (with 50Ω termination).Steps for
setting DC offset voltage to -150mV:
1. Press Menu Wave Param Offset
2. Use number keys to input the required number of -150.
3. Select corresponding unit mV.
Note: if the current DC offset value is valid, use this value. Press offset again, the parameters
become high level (MAX) and low level (MIN).
d)Set square wave
Default duty ratio: 50% limited by minimum pulse width 6.5ns. Steps to set a square wave
with 1Hz, 1.5Vpp, DC offset 0V, duty ratio 70%:
1. Press Menu Wave Type Square Param
2. Enter required numerical value and select the unit.