3.3 Basic Waveform Output
This chapter introduces how to output different waveforms. Default waveform: sine wave
with 1 kHz and 100mVpp.
a) Set output frequency
Default waveform: A sine wave of 1 kHz frequency and 100mV amplitude (with 50Ω
termination).
Steps for setting the frequency to 2.5MHz:
1) Press Menu → Wave → Param → Freq in turn to set the frequency. Pressing Freq to
edit frequency or period.
2) Press the number keyboard to input 2.5.
3) Select corresponding unit on the bottom bezel menu: MHz
b)Set output amplitude
Default waveform: A sine wave of 100mV peak-peak value with 50Ω termination.
Steps for setting the amplitude to 300mV:
1. Press Menu → Wave → Param → Amp in turn. Press Amp again can switch between
Vpp, Vrms, and dBm.
2. Press number button to input 300.
3. Press unit mVpp.