User Manual

UTE9802 Programming manual
*IDN?
This query command reads the identification string of the instrument. Return value example
UTE9802
SN202111051001V1.12V0.02
MEAS:VOLT
Read the measured voltage value.
Return value example220.0VOLOut of
range, the same below
MEAS:CURR
Read the measured current value.
Return value example5.12A or 500.1mA
MEAS:POW
Read the measured power value. Return value example350.02W
MEAS:PF
Read the measured power factor value. Return value example0.501
MEAS:FREQ
Read the measured frequency value. Return value example50.01Hz
MEAS:ALL
Read back all measured values at one time: measurement mode, voltage,
current, power, power factor, frequency, separated by commas.
ExampleAC,220.1V5.01A1103.2W0.90149.99Hz
ExampleDC,220.1V5.01A1103.2W
ExampleRMS,220.1V5.01A1103.2W0.99849.99Hz
MEAS:MODE
Read the measurement mode. Return value exampleACDCRMSAC+DC
SET:MODE XXX
Set the measurement mode. Return value exampleSUCCESSERR
RANG:VOLT
Read the voltage range. Return value exampleAUTO
75V150V300V600V
RANG:VOLT XXXX
Set the voltage range. Return value exampleSUCCERR
RANG:CURR
Read the current range. Return value exampleAUTO 500mA2A8A20A
RANG:CURR XXXX