User Manual

connected to EXT TRIG interface. Signal can be
used for EXT and EXT can be used within signal
trigger level scope of -3V to +3V
.
A
utomatic trigger: The system will acquire
waveform data automatically so as to displa
y
scanning base line on the screen without trigger
signal input. It will convert to trigger scannin
g
automatically for signal synchronization whe
n
producing trigger signal
.
It is allowed to select EXT/5 for oscillograph of 5.7-
inch screen. Outer trigger shall be divided by 5 so
that trigger scope can expand to -15V to +15V for
trigger of digital storage oscillograph under large
signal
.
Municipal power: Such trigger mode can be used
to observe signal with regard to municipal power
.
For instance, it can provide relationship betwee
n
lighting equipment and power supply equipment so
as to gain stable synchronization
.
2 Trigger mode: It is to determine behavior mode
of digital storage oscillograph without trigger event
.
This digital storage oscillograph can also provide
3
triggers, namely, automatic, normal and single
trigger modes
.
Ƶ
Ƶ
Operating Manual for UTD2000/3000
4
2
Note: Without trigger signal is allowed under
automatic” m ode when setting scanning waveform to
5-ms/div or less time base
.
Normal trigger: Digital storage oscillograph ca
n
not acquire waveform until meeting trigger conditions
under ordinary trigger mode. Tr igger scanning can be
produced when producing trigger signal
.
Single trigger: User can press “operation” button
once under single trigger mode for entry and trigger
waiting for digital storage oscillograph. It will sample
and display all acquired waveform then stop after
digital storage oscillograph detects once trigger
.
Ƶ
Ƶ